MedVue Firmware Update

To download files: Right-click on the file name, then select “Save Target As”.

Revision 1.01.08 - Effective May 8, 2014

Rev 1.01.08 - Fixed issue with Height/BMI locking with weight when lock key pressed or auto-lock occurred.

Rev 1.01.07 - Update to fix production issue with MV2.  Update for new MVWIFI option card hardware.

Rev 1.01.06 - Update MedVue application software to fix an issue with the Ethernet/WiFi option card.  Adjust display contrast to help with the new white backlight.  Improvements to help extend battery life.

Rev 1.01.05 - Update MedVue application software adding OIML Yes/No prompt to prepare for MV2 production.

Rev 1.01.04 - Update MedVue application software to fix a bug when communicating with Welch Allyn devices.

Rev 1.01.03 - Update MedVue application software to fix a bug when option card installed that would affect the display.

Rev 1.01.02 - Update MedVue application software to fix a bug in detecting the SDCard.

Rev 1.00.01 - Update MedVue application software to fix a bug in serial/TCP/USB communication.

Rev 1.00.00 - Initial software release.

Software Update Instructions:

In order to update the MedVue indicator you will need a USB-A to USB-B cable, a Phillips head screwdriver, and a paper clip.

Software Update Procedure:

• Download the current MedVue software.
• Unzip the downloaded file by right clicking and select "Extract All".
• Using the Phillips head screwdriver, remove the left end cap of the indicator.
• Using the paper clip, press the calibration switch and power on the indicator.
• The indicator should now display "boot rev 1.0" at the bottom of the display.
• Connect the indicator to your PC using the USB-A to USB-B cable.
• You should see a new removable storage device appear in My Computer.
• Find the unzipped file from step 2 above, it should be named "MedVue.afx.S19".
• Drag and drop the file to the new removable storage drive named "MedVue".
• The indicator display should show "Flashing S19" while the update is progressing.
• Once done, the indicator should reset and the new revision number should appear during startup.
• If you have difficulty, try starting over from the beginning of these steps.

DETECTO   •   Ph. (800) 641-2008   •   detecto@cardet.com   •   203 East Daugherty St., Webb City, MO 64870

Please Login to View Your Account



Create an Account


Click here to register.


Forgot Your Username or Password?


Click here to retrieve your account details.

Forgot Your Password?

Please enter your username or email address. You will receive a link to create a new password via email.



Quick Quote

  • This field is for validation purposes and should be left unchanged.